General Attributes:
Lakeshore Model 480 fluxmeter very good price
The Model 480 fluxmeter is a high-performance instrument designed for use in industrial and measurement system environments. It measures total magnetic flux, which can then be used to calculate flux density (B) and magnetic field strength (H). This device is ideal for a variety of applications, including magnetizing processes, manual and automated magnet testing and sorting, and serving as the core component in BH loop or hysteresis measurement systems. The Model 480 is also compatible with most sensing coils and fixtures, making it a versatile tool for magnetic measurement tasks.
- 5¾ digit DC resolution (1 part out of ±300,000)
- Automatic drift compensation
- Very fast peak capture
- AC frequency response to 50 kHz
- IEEE-488 and serial interfaces
- Storage of parameters for up to 10 existing coils
480 fluxmeter specifications
Measurement
Number of units: 1
Input type: 2-lead, ground-referenced
Input resistance: 100 kΩ or 10 kΩ
Maximum operating input voltage: 60 V
Absolute maximum input voltage: 100 V — WARNING — voltages between 60 V and 100 V will not damage the instrument but could result in personal injury or damage to other instruments
Update rate: 5 rdg/s display; 30 rdg/s IEEE-488; 30 rdg/s serial
DC
DC resolution: To 5¾ digits
DC integrator capacitance: 1 µF nominal
DC input resistance: 100 kΩ 10 kΩ
DC ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs
DC resolution: 0.001 mVs, 0.0005 mVs, 0.0005 mVs, 0.0005 mVs
DC accuracy: Offset: ±10 µVs ±DC integrator drift; Gain: ±0.25% of reading (<10 Vs/s maximum rate of change)
DC minimum dΦ/dt: 20 µVs/min
DC maximum dΦ/dt: 60 Vs/s
DC integrator drift: ±1 µVs/min, 0.0004% FS/min on 300 mVs range (100 kΩ input resistance constant temperature environment)
DC peak
DC peak resolution: 4¾ digits
DC peak integrator capacitance: 1 µF nominal
DC peak input resistance: 100 kΩ 10 kΩ
DC peak ranges: 300 mVs, 30 mVs, 30 mVs, 3 mVs
DC peak resolution: 0.01 mVs, 0.001 mVs, 0.001 mVs, 0.001 mVs
DC peak minimum reading: 0.05 mVs, 0.005 mVs, 0.005 mVs, 0.005 mVs
DC peak accuracy: Offset: ±100 µVs ±DC integrator drift; Gain: ±5% of reading (<10 Vs/s maximum rate of change)
DC peak maximum dF/dt: 60 Vs/s
DC peak update rate: Reduces update rate to ¼ normal
AC
AC resolution: 4¾ digits (reduced to 3¾ digits on the 30 µVs range)
AC integrator capacitance: 0.1 µF nominal
AC input resistance: 100 kΩ
AC ranges: 30 mVs, 3 mVs, 300 µVs, 30 µVs
AC resolution: 0.001 mVs, 0.0001 mVs, 0.01 µVs, 0.01 µVs
AC minimum rdg: 3.000 mVs, 0.3000 mVs, 30.00 µVs, 3.00 µVs
AC frequency response: 2 Hz to 50 kHz
AC accuracy: ±1% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±5% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)
AC integrator drift: NA
AC peak
AC peak resolution: 3¾ digits
AC peak integrator capacitance: 0.1 µF nominal
AC input resistance: 100 kΩ
AC peak ranges: 30 mVs, 3 mVs, 300 µVs
AC peak resolution: 0.01 mVs, 0.001 mVs, 1 µVs
AC minimum reading: 0.01 mVs, 0.001 mVs, 5 µVs
AC peak accuracy: ±5% rdg ±10 µVs (10 Hz to 10 kHz sinusoidal); ±10% rdg ±10 µVs (2 Hz to 50 kHz sinusoidal)
AC peak update rate: Reduces update rate to ¼ normal
Front panel
Display type: 2-line × 20-character, vacuum fluorescent display
Display resolution: To ±5¾ digits
Display update rate: 5 rdg/s
Display units: Vs, MxN, WbN, VF, MxF, WbF, G, T, Wbcm, A, %
Units multipliers: p, n, µ, m, k, M, G
Annunciators: AC – AC input signal, DC – DC input signal, ∧ ∨ – Positive and negative peaks, R – Remote operation, :Alarm on
Keypad: 21 full travel keys
Interfaces
IEEE-488.2 capabilities: SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1
Serial interface: RS-232C electrical, DA-9 connector, 9600 BAUD
External reset type: Contact closure
Software support: LabVIEW™ driver
Alarms
Number: 2
Settings: High and low set point, Inside/Outside, Audible
Actuators: Display annunciator, beeper, relays
Relays
Number: 3
Contacts: Normally open (NO), normally closed (NC), and common (C)
Contact rating: 30 VDC at 2 A
Operation: Follow high, low alarms with third relay indicating no alarm state — can be operated manually
Connector: Detachable terminal block
Monitor analog output
Scale: ±3 V = ±FS on Vs range
Accuracy: ±1% of reading ±10 mV, (DC to 10 kHz); ±5% of reading ±10 mv, (10 kHz to 50 kHz)
Minimum load resistance: 1 kΩ
Connector: Detachable terminal block
Corrected analog output
Scale: User selected
Range: ±10 V
Resolution: 0.3 mV
Accuracy: ±2.5 mV
Minimum load resistance: 1 kΩ
Connector: Detachable terminal block/p>
General
Ambient temperature: 15 ºC to 35 ºC at rated accuracy; 5 ºC to 40 ºC with reduced accuracy
Power requirement: 100, 120, 220, 240 VAC, +5%, -10%, 50 or 60 Hz, 20 VA
Size: 216 mm W × 89 mm H × 318 mm D (8.5 in × 3.5 in × 12.5 in), half rack
Weight: 3 kg (6.6 lb)