ORTEC Model 567 Time-to-Amplitude Converter/Single-Channel Analyzer (TAC/SCA)

Category:

▶ ORTEC/AMETEK

keyWord:

ORTEC Model 567 Time-to-Amplitude Converter/Single-Channel Analyzer (TAC/SCA)

Describe:

567 Time-to-Amplitude Converter / Singe Channel Analyzer


Contact us

Description

http://www.jinstartech.com/

General Attributes:

ORTEC Model 567 very good price

 

  • For time spectroscopy in the range from 10 ns to 2 ms
  • Includes SCA to set a time window for coincidence experiments
  • Valid Start and Valid Conversion outputs
  • Selectable output delay and width
  • Output synchronized with a stop or external strobe signal
  • Provision to reject unwanted start or stop input signals
  • Positive or negative input signals


The ORTEC Model 567 Time-to-Amplitude Converter/Single-Channel Analyzer (TAC/SCA) measures the time interval between start and stop input pulses, generates an analog output pulse proportional to the measured time, and provides built-in single-channel analysis of the analog signal. Additional gating modules are not necessary with this unit, and timing experiments requiring time ranges of 10 ns to 2 ms may be performed with single-channel analysis, giving the experimenter unparalleled flexibility in analyzing random events that occur within a selected time range.

Separate gating (anticoincidence or coincidence) of the start and stop inputs eliminates unwanted events from the time spectra via externally imposed energy or timing restrictions. The Model 567 also incorporates a built-in SCA inhibit feature in which a TAC output is available only if the output pulse falls within the window restrictions imposed by the SCA. This feature may be switched in or out by a convenient front-panel switch.

In addition to its start and stop input gating capabilities, the Model 567 provides for a pulsed or dc-level Reset/Inhibit signal via a front-panel input connector. A Reset/Inhibit input signal terminates the conversion cycle and maintains a reset condition, inhibiting further TAC conversions for the duration of the Reset/Inhibit pulse. A TAC output pulse that is in process at the time a Reset/Inhibit input is received will be completed before converter reset is initiated.

Valid Start and Valid Conversion outputs are provided for each accepted start and stop input respectively. The duration of the Valid Start output indicates the interval from the accepted start until the end of reset. Valid Conversion occurs from the end of the internal delay after stop to the end of reset.

The selectable TAC output width and variable delay, which are easily adjustable, further serve to make the Model 567 a flexible instrument. The output of the TAC may be synchronized with the stop signal or an external strobe signal to further enhance its versatility.

The single-channel analyzer section of the Model 567 allows the experimenter to place very specific time restrictions on the timing spectrum. The SCA is operated in the Window mode, where the upper-level discriminator setting is added to that of the lower-level discriminator. The SCA output pulse width is equal to the time from the occurrence of the TAC output until the end of the reset pulse or the end of the TAC output. The synchronization of the SCA output with the stop input virtually eliminates any time walk in the SCA output.

All Model 567 inputs are printed wiring board (PWB) jumper-selectable to accept either negative or positive NIM standard signals. All inputs and outputs are dc-coupled so that changing input count rates will not hinder normal operation of the Model 567. The TAC output should be connected to the dc-coupled input of a multichannel analyzer (MCA) for optimum high count-rate performance.